Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Liquiline System CA80AL - High-precision process analyzer for aluminum measurement

Aluminum analyzer
Liquiline System CA80AL

Colorimetric system for drinking water and wastewater monitoring and optimization of process control

CA80AL
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    Colorimetric

  • Characteristic

    Process analyzer for aluminum in aqueous solutions
    Comply with standard colorimetric measuring principle - pyrocatechol violet methot

  • Size

    Housing (open version):
    793 x 530 x 417 mm
    31.22 x 20.87 x 16.42 in
    Housing (closed version):
    793 x 530 x 463 mm
    31.22 x 20.87 x 18.23 in
    Housing with base:
    1723 x 530 x 463 mm
    67.83 x 20.87 x 18.23 in

  • Design

    Open design, cabinet and stand housing
    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 ... 40 °C (39.2 ... 104 °F)

  • Ambient temperature

    5 ... 40C ( 41 ... 104F)
    Outdoor version: - 20 to 40 °C (-4 to 104 °F)

  • Process pressure

    at atmospheric pressure, < 0.2 bar

  • Sample flow rate

    min. 5ml/min (0,17 fl.oz/min)

  • Consistency of the sample

    low solid content (< 50NTU), aqueous homogeneous

  • Application

    Monitoring of the aluminum content of potable water, mineral water and industrial waste water

  • Power supply

    100 ... 120 VAC / 200 ... 240 VAC ± 10%
    24 VDC ± 10%
    50 ± 1 oder 60 ± 1,2 Hz

  • Output / communication

    2x 0/4 ... 20 mA
    Webserver, Modbus, Ethernet/IP, Profibus DP

  • Input

    1 or 2 measuring channel
    1 ... 4 digital sensor inputs for sensors with Memosens protocol (optional)

  • Measuring range

    15 to 1000 µg/l Al
    15 to 1000 µg/l with dilution function to maximum 300 to 20 000 µg/l Al

  • Consumables

    Reagents and standard solutions CY80AL as well as cleaner CY800 are necessary for the operation. Regular maintenance is done with the parts of the maintenance kit CAV800.