Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Liquiline Control CDC90 is an automatic cleaning and calibration system for pH and ORP sensors.

Cleaning and calibration system
Liquiline Control CDC90

Fully automatic system for pH and ORP measuring points in all industries

CDC90
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    Potentiometric

  • Application

    Increased occupational safety by minimizing maintenance and servicing measures.
    Customizable cleaning and calibration cycles for up to two measuring points.
    Minimization of operating costs and buffer and cleaning agent consumption through smart media distribution and dosing.

  • Installation

    Complete automated cleaning and validation system including a selectable retractable holder

  • Characteristic

    Fully assembled system with powder-coated or stainless steel pneumatic control unit.
    All non-ex applications

  • Measuring principle

    Automated cleaning and calibration of Memosens pH and ORP sensors installed in retractable assemblies.

  • Design

    Complete cleaning and calibration system including dosing pumps and canisters for buffer and cleaning solutions for connection to various retractable assemblies and sensors

  • Dimension

    268 x 1170 x 770 mm (l x h x w)
    10.55 x 46.06 x 28.74 inch (l x h x w)

  • Connection

    DCS/PLS connection on Modbus TCP, Ethernet IP, Profinet, ProfibusDP and current in/ output

  • Input

    1 to 2x Memosens digital sensor
    1x 0/4 to 20mA input
    4x digital input

  • Output / communication

    DCS/PLS connection via Modbus TCP, EtherNet/IP, PROFINET, PROFIBUS DP, digital and analog signals

  • Additional certifications

    UKCA, CSA general purpose