  • Measuring principle

    Total chlorine

  • Application

    Wastewater treatment plants, disinfection monitoring and control in wastewater outlet, reuse of outlet water, utilities of all industries, all disinfection measuring points

  • Characteristic

    Amperometric measurement of total chlorine

  • Measuring range

    0.1 to 10 mg/l total chlorine

  • Measuring method

    Total chlorine consists of free chlorine (HOCl, OCl-) and bound chlorine (chloramines)
    All components are reduced at the working electrode
    Almost pH independent

  • Design

    Closed, membrane covered measuring cell

  • Material

    Sensor shaft: PVC
    Membrane: PET
    Membrane cap: PPE

  • Dimension

    Diameter: 25 mm (0.98 inch)
    Length: 148 mm (5.83 inch)

  • Process temperature

    5 to 45 °C
    (41 to 113 °F)

  • Process pressure

    1 to 4 bar (14.5 to 58 psi)

  • Temperature sensor

    10k NTC integrated (Memosens)

  • Connection

    Inductive, digital connection head with Memosens