  • Measuring principle

    Transmittance measurement

  • Measured variables

    Transmittance, opacity, relative opacity, extinction, dust concentration

  • Measuring range

    Transmittance :100 ... 50 % / 100 ... 0 %
    Opacity: 0 ... + 50 % / 0 ... 100 %
    Relative opacity: 0 ... + 50 % / 0 ... 100 %
    Extinction: 0 ... + 0.3 / 0 ... 1

    Dust concentration: 0 ... + 200 mg/m³ / 0 ... 10,000 mg/m³
    The measurement depends on measuring distance and dust properties

  • Process temperature

    –40 °C ... +600 °C

  • Process pressure

    With MCU-P control unit: –50 hPa ... 2 hPa With external purge air unit: –50 hPa ... 30 hPa

  • Degree of protection

    IP66, IP54

  • Product safety

    CE