Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Product picture SS2100 TDLAS gas analyzer, H2O, CO2, H2S, NH3, C2H2 measurements, right angle view

SS2100 gas analyzer

Exceptionally reliable for measuring trace gas components

SS2100
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    TDLAS

  • Product Headline

    An advanced gas analyzer for hydrogen sulfide (H2S) measurement. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.

  • Channels

    1

  • Analyte and Measurement ranges

    H2S (Hydrogen Sulfide):
    0 to 10 ppmv
    0 to 1000 ppmv
    0 to 5000 ppmv
    0 to 5% volume

  • Measured Variables

    Concentration
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W
    (18 to 24 VDC - optional for electronics only)
    3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz

  • Communication

    Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C and Ethernet
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs: 2, Flow Alarm, Validation Request

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    IP66, Type 4X

  • Measuring principle

    TDLAS

  • Product Headline

    An advanced gas analyzer for moisture (H2O) measurement. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.

  • Channels

    1

  • Analyte and Measurement ranges

    H2O (Moisture):
    0 to 10 ppmv
    0 to 100 ppmv
    0 to 50 ppmv
    0 to 5000 ppmv

  • Measured Variables

    Concentration
    Moisture Dew Point
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W
    (18 to 24 VDC - optional for electronics only)
    3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz

  • Communication

    Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C and Ethernet
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs: 2, Flow Alarm, Validation Request

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    IP66, Type 4X

  • Measuring principle

    TDLAS

  • Product Headline

    An advanced gas analyzer for carbon dioxide (CO2) measurement. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.

  • Channels

    1

  • Analyte and Measurement ranges

    CO2 (Carbon Dioxide):
    0 to 10 ppmv
    0 to 1000 ppmv
    0 ppmv to 5000 ppmv
    0 to 5% volume
    0 to 10% volume
    0 to 20% volume

  • Measured Variables

    Concentration
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W
    (18 to 24 VDC - optional for electronics only)
    3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz

  • Communication

    Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C and Ethernet
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs: 2, Flow Alarm, Validation Request

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    IP66, Type 4X

  • Measuring principle

    TDLAS

  • Product Headline

    An advanced gas analyzer for ammonia (NH3) measurement. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.

  • Channels

    1

  • Analyte and Measurement ranges

    NH3 (Ammonia): 0 to 5 ppmv

  • Measured Variables

    Concentration
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ±10%, 50 to 60 Hz, 300W
    (18 to 24 VDC - optional for electronics only)
    3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz

  • Communication

    Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C and Ethernet
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs: 2, Flow Alarm, Validation Request

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    IP66, Type 4X

  • Measuring principle

    TDLAS

  • Product Headline

    An advanced gas analyzer for acetylene (C2H2) measurement. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.

  • Channels

    1

  • Analyte and Measurement ranges

    C2H2 (Acetylene):
    0 to 5 ppmv (minimum)
    0 ppmv to 3000 ppmv (maximum)

  • Measured Variables

    Concentration
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W
    (18 to 24 VDC - optional for electronics only)
    3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz

  • Communication

    Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C and Ethernet
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs: 2, Flow Alarm, Validation Request

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    IP66, Type 4X

  • Measuring principle

    TDLAS

  • Product Headline

    Advanced multi-channel gas analyzer for hydrogen sulfide (H2S) and moisture (H2O), also known as a 2-Pack. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.

  • Channels

    2

  • Analyte and Measurement ranges

    H2S (Hydrogen Sulfide):
    0 to 10 ppmv (minimum)
    0 to 1000 ppmv (maximum)
    H2O (Moisture):
    0 to 100 ppmv (minimum)
    0 to 5000 ppmv (maximum)

  • Measured Variables

    Concentration
    Moisture Dew Point
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W
    (18 to 24 VDC - optional for electronics only)
    3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz

  • Communication

    Analog Output: Four 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (H2O)
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs (H2S): 2, Flow Alarm, Validation Request
    Digital Outputs (H2O): 2, Concentration Alarm, General Fault

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    IP66, Type 4X

  • Measuring principle

    TDLAS

  • Product Headline

    Advanced multi-channel gas analyzer for hydrogen sulfide (H2S) and carbon dioxide (CO2), also known as a 2-Pack. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.

  • Channels

    2

  • Analyte and Measurement ranges

    H2S (Hydrogen Sulfide):
    0 to 10 ppmv (minimum)
    0 to 1000 ppmv (maximum)
    CO2 (Carbon Dioxide):
    0 to 5000 ppmv
    0 to 5% volume
    0 to 10% volume
    0 to 20% volume

  • Measured Variables

    Concentration
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W
    (18 to 24 VDC - optional for electronics only)
    3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz

  • Communication

    Analog Output: Four 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (CO2)
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs (H2S): 2, Flow Alarm, Validation Request
    Digital Outputs (CO2): 2, Concentration Alarm, General Fault

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    IP66, Type 4X

  • Measuring principle

    TDLAS

  • Product Headline

    Advanced multi-channel gas analyzer for hydrogen sulfide (H2S), moisture (H2O), and carbon dioxide (CO2), also known as a 3-Pack. The SS2100 uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.

  • Channels

    3

  • Analyte and Measurement ranges

    H2S (Hydrogen Sulfide):
    0 to 10 ppmv (minimum)
    0 to 1000 ppmv (maximum)
    H2O (Moisture):
    0 to 100 ppmv (minimum)
    0 to 5000 ppmv (maximum)
    CO2 (Carbon Dioxide):
    0 to 5000 ppmv
    0 to 5% volume
    0 to 10% volume
    0 to 20% volume

  • Measured Variables

    Concentration
    Moisture Dew Point
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ± 10 %, 50 to 60 Hz, 300W
    (18 to 24 VDC - optional for electronics only)
    3 Amps max at 120 VAC , 1.5 Amps max at 240 VAC Hz

  • Communication

    Analog Output: Four 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (H2O/CO2)
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs (H2S): 2, Flow Alarm, Validation Request
    Digital Outputs (H2O/CO2): 2, Concentration Alarm, General Fault

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    IP66, Type 4X