Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Product picture protection tube TT411

iTHERM ModuLine TT411
Welded thermowell

Metric thermowell for hygienic and aseptic applications in the food & beverages and pharmaceutical industries

TT411
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    Fabricated Thermowell

  • Characteristic / Application

    metric style
    hygienic/aseptic design (3-A®, EHEDG, ASME BPE, FDA)
    hygienic process connections
    iTHERM QuickNeck for easy and cost saving recalibration

  • Head connection

    iTHERM QuickNeck
    male thread:
    M24x1,5
    G3/8"

  • Maximum standard immersion length

    4.000 mm (157,48")

  • Max. immersion length on request

    Not defined

  • Process connection

    compression fitting
    weld in adaptor
    clamp connections acc. to ISO2852
    screwed pipe joint acc. to DIN 11851
    aseptic screwed pipe joint acc. to DIN 11864-1
    metallic sealing system
    thread acc. to ISO 228 for liquiphant adaptor APV Inline
    Varivent
    Ingold
    SMS 1147
    Neumo Biocontrol
    Tee and elbow thermowells DIN11865

  • Thermowell root diameter

    6,0 mm (0,24")
    9,0 mm (0,35")
    12,7 mm (0,5")

  • Medium contact material

    1.4435 delta-ferrit < 1%
    316L

  • Wetted part finishing (Ra)

    < 0,76 μm (29,92 μin)
    < 0,38 μm (14,96 μin)
    < 0,38 μm (14,96 μin) electropolished

  • Tip shape

    straight
    reduced
    tappered

  • Temperature range

    -200...650 °C (-328...1.202 °F)

  • Max. process pressure (static)

    40 bar (580 psi)

  • Max. process pressure at 400 °C

    depends on process connection